==Microarray Fabrication== ===Casting=== # In a dish, mix 45 g silicone elastomer with 4.5 g PDMS curing agent (10:1 silicone:curing agent). # Mix well with plastic knife until bubbles form, approximately 1 min. # Place dish into vacuum chamber and close lid. # Ensure that all valves are in the correct position. (NEED FIGURE) # Turn on pump and wait for bubbles to rise to the surface of the mixture. # Close the small plastic valve on the vacuum chamber and shut off pump. If the pump runs for too long, the rising bubbles will cause the mixture to overflow the dish. # Allow the dish to seat in the sealed vacuum chamber for 20 minutes. Do not worry if there are still bubbles remaining. # Open the small plastic valve on the vacuum chamber. The rapid change in pressure should cause the rest of the bubbles to burst. Open the release valve on the tubing to fully equilibrate. # Pour the mixture over the silicon mold. # Vacuum again, 10 minutes. The pump does not need to be shut off this time as there is no risk of overflow. Do not worry if there are still some bubbles in the material. These bubbles will not affect the wells because the plate is being cast upside down; bubbles are rising to the eventual bottom of the plate. # Bake for 3 hours at 60 deg C. # Use a scalpel to cut around the edge of the dish first, then around the edge of the silicon mold. Peel away and discard this ring. # Use the scalpel to peel the edge of the PDMS disk away from the silicon mold. Carefully work your way around the entire disk. # Carefully peel the PDMS disk away from the silicon mold and place in a new dish. Cut along the edges of the plates. Each disk yields 2 plates. # Transport the excised plates in a covered dish with parafilm around the edge to keep out dust. ===Mounting=== # Use red, 2mm (height) rubber gasket, with 16 squares. Peel off white backing to expose glue. # Place gasket sticky side up on hood shelf, along with PDMS plate. UV for 10-15 minutes. # Attach gasket to PDMS, aligning the edges. Trim any excess PDMS (e.g., due to sloppy cutting after casting). # Clean glass slide and mount PDMS/gasket assembly, gasket side up. The PDMS will stick to the glass. # Cover with scotch tape to keep out dust.
Summary:
Please note that all contributions to ZhangLabWiki may be edited, altered, or removed by other contributors. If you do not want your writing to be edited mercilessly, then do not submit it here.
You are also promising us that you wrote this yourself, or copied it from a public domain or similar free resource (see
ZhangLabWiki:Copyrights
for details).
Do not submit copyrighted work without permission!
Cancel
Editing help
(opens in new window)
Navigation menu
Personal tools
Not logged in
Talk
Contributions
Create account
Log in
Namespaces
Page
Discussion
English
Views
Read
Edit
View history
More
Search
Navigation
Main Page
Current events
Recent changes
Random page
Investigators
Matt Cai
Song Chen
Eric Chu
Dinh Diep
Elizabeth Duong
Shicheng Guo
Alan Fung
Daniel Jacobsen
Blue Lake
Huy Lam
Alice Li
Andrew Richards
Brandon Sos
Chris Wei
Yan Wu
Kun Zhang
Tools
What links here
Related changes
Special pages
Page information