Leica SP8 Confocal Microscope: Difference between revisions

From ZhangLabWiki
Jump to navigation Jump to search
>Hosuki78
No edit summary
>Mzcai
 
(17 intermediate revisions by one other user not shown)
Line 2: Line 2:
PFBH 036
PFBH 036


{| {{table}}
| [[File:Confocal01.jpg|400px]]
||
| [[File:Confocal02.jpg|400px]]
|}


===Leica support contact===
===Leica support contact===
*Technical support: (Toll Free) 866-830-0735
#Technical support: (Toll Free) 866-830-0735
 
#Clark Thom
*Clark Thom
#*West Coast Team Leader Field Service
West Coast Team Leader Field Service
#*619-278-8324
619-278-8324
#*clark.thom@leica-microsystems.com
clark.thom@leica-microsystems.com
#Carlos Alonso
#*Confocal Applications Specialist
#*858-353-8062
#*carlos.alonso@leica-microsystems.com


*Carlos Alonso
*(Call technical support first before trying to contact Clark)
Confocal Applications Specialist
858-353-8062
carlos.alonso@leica-microsystems.com




Line 39: Line 34:
*Only one important consideration is that substrate must be '''#1.5 glass slide'''!
*Only one important consideration is that substrate must be '''#1.5 glass slide'''!


===Starting and Finishing sequence===
 
===Workstation===
*(This workstation has dynamic IP address, so the IP information below can be changed)
*IP address : 132.239.236.163
*Subnet Mask : 255.255.255.0
*Gateway : 132.239.236.1
*DNS : 132.239.1.19
*MAC : B4:B5:2F:CC:91:7F
*Network Port: B.1.32 (Room #038)
*Login:
**admin: !Admin12345
**ConfocalUser: True!Confocal55
 
===Turning on/off sequence===
#Turn on sequence
#*Confocal Power supply : '''Turn on from left to right''' Scan power --> Laser power --> Key (turn clockwise)
#*PC power strip : Turn on the switch on the strip
#Turn off sequence --> Reverse way of turn on sequence
#*Shut down PC
#*Turn off PC power strip
#*Turn off confocal power supply : '''Turn off from right to left''' : Key --> Laser power --> Scan power
 
 
{| {{table}}
| [[File:ConfocalPower_TurnOn.jpg|250px]]
||
| [[File:ConfocalPower_TurnOff.jpg|250px]]
|}

Latest revision as of 20:49, 18 November 2016

Location[edit]

PFBH 036


Leica support contact[edit]

  1. Technical support: (Toll Free) 866-830-0735
  2. Clark Thom
    • West Coast Team Leader Field Service
    • 619-278-8324
    • clark.thom@leica-microsystems.com
  3. Carlos Alonso
    • Confocal Applications Specialist
    • 858-353-8062
    • carlos.alonso@leica-microsystems.com
  • (Call technical support first before trying to contact Clark)


General specification[edit]

  1. 3 Laser source – 488, 552, 638nm, Solid State, up to 20mW
  2. 2 PMT (Photo Multiplier Tube)
  3. 3 Objective Lens
    • 10x NA = 0.3
    • 20x NA = 0.75(Glycerine or Oil immersion)
    • 63x NA=1.4(Oil immersion)
  4. Z-axis step size – 50nm
  5. Scan area - 1.16 mm x 1.16mm (@10x obj., zoom = 1x) ~ 24.22 um x 24.22um (@10x obj., zoom = 48x)
  6. Image size - up to 8192 x 8192 pixel
    • 1 pixel = 141.92nm @10x obj., 22.53nm @ 63x obj.
    • Max. Scan field : 21.2mm (Tile scanning)


Sample preparation[edit]

  • Only one important consideration is that substrate must be #1.5 glass slide!


Workstation[edit]

  • (This workstation has dynamic IP address, so the IP information below can be changed)
  • IP address : 132.239.236.163
  • Subnet Mask : 255.255.255.0
  • Gateway : 132.239.236.1
  • DNS : 132.239.1.19
  • MAC : B4:B5:2F:CC:91:7F
  • Network Port: B.1.32 (Room #038)
  • Login:
    • admin: !Admin12345
    • ConfocalUser: True!Confocal55

Turning on/off sequence[edit]

  1. Turn on sequence
    • Confocal Power supply : Turn on from left to right Scan power --> Laser power --> Key (turn clockwise)
    • PC power strip : Turn on the switch on the strip
  2. Turn off sequence --> Reverse way of turn on sequence
    • Shut down PC
    • Turn off PC power strip
    • Turn off confocal power supply : Turn off from right to left : Key --> Laser power --> Scan power


File:ConfocalPower TurnOn.jpg File:ConfocalPower TurnOff.jpg