Matt:LabNotes/2014-2-25: Difference between revisions
Jump to navigation
Jump to search
>Mzcai |
>Mzcai |
||
Line 67: | Line 67: | ||
====Pre-circularized Padlock Probes (C7 & C8)==== | ====Pre-circularized Padlock Probes (C7 & C8)==== | ||
*Combined all the pre-circle ppMALAT1 made on: [[Matt:LabNotes/2014-2-10]] into | *Combined all the pre-circle ppMALAT1 made on: [[Matt:LabNotes/2014-2-10]] into 64ul | ||
*75% EtOH for 10 minutes | *75% EtOH for 10 minutes | ||
*Wash with 1X PBS 2 times | *Wash with 1X PBS 2 times | ||
*0.1N HCl for 2 minutes | *0.1N HCl for 2 minutes | ||
*Wash with 1X PBS 3 times | *Wash with 1X PBS 3 times | ||
*Add | *Add 32ul to C7 & C8 | ||
====Fixed Pre-circularized Padlock Probes (F7 & F8)==== | ====Fixed Pre-circularized Padlock Probes (F7 & F8)==== | ||
*Not going to do anything during Ampligase step | *Not going to do anything during Ampligase step |
Revision as of 01:12, 26 February 2014
Secondary Rolony Generation
Samples: Matt:LabNotes/2014-2-21
Procedure
Anneal Padlock Probes
- 1X Ampligase Buffer
- 2uM of each padlock probe
- 0.5 U/ul Ampligase
- Incubate @ 45C for 24 hours
Fixed Template - 25U Ampligase (C9 & F9)
Component | Volume |
10X Ampligase Buffer | 10 |
200uM ppMALAT1 | 1 |
5U/ul Ampligase | 10 |
H2O | 79 |
Total | 100 |
Fixed Template - 50U Ampligase (C10 & F10)
Component | Volume |
10X Ampligase Buffer | 10 |
200uM ppMALAT1 | 1 |
5U/ul Ampligase | 20 |
H2O | 69 |
Total | 100 |
1st Rolony (E6 & F6)
Component | Volume |
10X Ampligase Buffer | 10 |
200uM ppMALAT1 | 1 |
200uM ppACTB | 1 |
200uM ppRAB7A | 1 |
5U/ul Ampligase | 10 |
H2O | 77 |
Total | 100 |
Empty Negative Control (D7 & E7)
- Not going to waste Ampligase since it's highly unlikely anything will be ligated, but will do RCA in these wells
Pre-circularized Padlock Probes (C7 & C8)
- Combined all the pre-circle ppMALAT1 made on: Matt:LabNotes/2014-2-10 into 64ul
- 75% EtOH for 10 minutes
- Wash with 1X PBS 2 times
- 0.1N HCl for 2 minutes
- Wash with 1X PBS 3 times
- Add 32ul to C7 & C8
Fixed Pre-circularized Padlock Probes (F7 & F8)
- Not going to do anything during Ampligase step