MDA with PGP1F, 6th
- MDA run for 8hr @06/27
- 400mM ALS --> original protocol
- Slow mixing --> 1 sec period, mixing for 5min
- Run MDA in ring mixer --> 25th chamber!
- Push ring mixer volume with air
- Left 8 chamber were empty, too many empty chamber...
Result
- qPCR result looked good.
- Most of chambers have 500pg~2400pg of amplicons.
- Left 8 chamber group were nearly empty, so that's why ch6 has almost no amplicon was detected.
- 25th chamber(Ring mixer) had less amount than other chamber did, but there was amplicon!
- It seemed that low conc. of ALS doesn’t work in the MDA device… I’ll try higher conc. than 133mM, such as 250mM ALS.
File:QPCR Result 2013-06-28 Edited.png File:QPCR Result MassTable 2013-06-28.png
CoRE fragmentation
- Done @06/28