AlanFung:Protocol/OliGreen ssDNA Stain: Difference between revisions

From ZhangLabWiki
Jump to navigation Jump to search
>Alan6017518
>Alan6017518
Line 23: Line 23:
*Determine the volume of Working Reagent Required
*Determine the volume of Working Reagent Required
  (# standards (5) + # samples) x (replicates (2)) x (volume per well (100 μL))
  (# standards (5) + # samples) x (replicates (2)) x (volume per well (100 μL))
*Prepare an aqueous working solution of the Quant-iT OliGreen reagent by making a 200 fold dilution of the concentrated DMSO solution in 1XTE Buffer
*Add 437.8uL of 1X TE Buffer to the 2.2uL dye aliquot


==Step 2: Prepare Oligonucleotide Standard==
==Step 2: Prepare Oligonucleotide Standard==

Revision as of 15:45, 31 May 2009

Probe Preparation

Samples & Materials

  • Quant-iT OliGreen ss DNA Assay Kit (Cat. no. O11492)
*Quant-iT OliGreen ss DNA Reagent (Component A), solution indimethylsulfoxide (DMSO)
*20X TE Buffer (Component B)
*Oligonucleotide Standard (Component C) 18-base M13 sequence primer, 
 with the sequence 5'-TGTAAAACGACGGCCAGT-3'
  • Sterile, distilled, DNase-free water

Overview

Procedures

Step 1: Preparing Working Reagent

*First Time
  • Allow the reagent to warm to RT before opening the vial
  • Aliquot reagent dye to 0.5mLtubes (2.2uL each tube) - Stor at -20C
*First Time
  • Dilute 20X TE buffer to 1X working solution
*Take 1mL 20X TE buffer and dilute it with 19mL DNase-free water
  • Determine the volume of Working Reagent Required
(# standards (5) + # samples) x (replicates (2)) x (volume per well (100 μL))

Step 2: Prepare Oligonucleotide Standard

  • Dilute the oligonucleotide standard provided(100ug/mL) to 100ng/mL with 1XTE buffer

File:OliGreen Standard.jpg

  • Load both samples and standard and into strip tubes
  • The strip tubes should look like this:

File:96 well plate.jpg

  • with the letters A-E (in A1-B2) being the standards and the numbers 1-11 (in C2-H4) being the samples
  • (it is read top-to-bottom then left-to-right; A1 – H1, A2 - H2, etc)
  • Mix well and incubate for 5 mins at RT, protected from light
  • After incubation, measure the sample fluorescence using the RT-PCR
  • Standard fluorescein wavelength (excitation ~480nm, emmision ~520nm)
  • Measure the absorbance at 520nm on the RT-PCR
  • For the Chromo4 Real Time PCR
  • Open the program
  • Subtract the fluorescnece value of the reagent blank from that of each of the samples
  • Use corrected data to generate a standard curve of fluorescence versus oligonucleotide concentration.
  • Record the ssDNA concentrations calculated by the worksheet

Step 3: Prepare Sample

  • Dilute sample to range of 50pg/mL to 50ng/mL
  • Typical probes concentration ranges from 4-20ng/uL
  • Perform a dilute according to the Table

Step 4: Prepare OliGreen Reagent

  • Prepare an aqueous working solution of the Quant-iT OliGreen reagent by making a 200 fold dilution of the concentrated DMSO solution in 1XTE Buffer
  • Add 437.8uL of 1X TE Buffer to the 2.2uL dye aliquot

Step 3: Sample Anaylysis

  • Dilute your sample solution in TE to a final volume of 1mL
  • Add 200uL of the aqueous working solution of the Quant-T OliGreen reagent to each sample
  • Incubate for 5 mins at room temperature, protected from light
  • Measure the fluorescence of the sample
  • Subtract the fluorescence value of the reagent blank from that of each of the samples
  • Determine the oligonucleotide concentration of the sample from the standard curve generated in Oligonucleotide Standard Curve