Hosuk:LabNotesMDA/2013-6-28: Difference between revisions
Jump to navigation
Jump to search
>Hosuki78 (→Result) |
>Hosuki78 (→Result) |
||
Line 17: | Line 17: | ||
*Left 8 chamber group were nearly empty, so that's why ch6 has almost no amplicon was detected. | *Left 8 chamber group were nearly empty, so that's why ch6 has almost no amplicon was detected. | ||
*25th chamber(Ring mixer) had less amount than other chamber did, but there was amplicon! | *25th chamber(Ring mixer) had less amount than other chamber did, but there was amplicon! | ||
*It seemed that low conc. of ALS doesn’t work in the MDA | *It seemed that low conc. of ALS doesn’t work in the MDA device. | ||
[[File:qPCR_Result_2013-06-28_Edited.png|800px]] [[File:qPCR_Result_MassTable_2013-06-28.png|250px]] | [[File:qPCR_Result_2013-06-28_Edited.png|800px]] [[File:qPCR_Result_MassTable_2013-06-28.png|250px]] | ||
====Considerations==== | |||
**In 20uL reaction, ALS is used 1.5uL for 1uL of cell, so 1.5/2.5 x 400mM KOH = 240mM in ALS+Cell while denaturing. | |||
*Therefore current 400mM of KOH in ALS is much higher concentration for MDA device, because KOH conc. in ring mixer while denaturing is 357mM. | |||
*Let’s try 240mM in ring mixer! | |||
====CoRE fragmentation==== | ====CoRE fragmentation==== | ||
*Done @06/28 | *Done @06/28 |
Revision as of 18:08, 28 June 2013
MDA with PGP1F, 6th
- MDA run for 8hr @06/27
- 400mM ALS --> original protocol
- Slow mixing --> 1 sec period, mixing for 5min
- Run MDA in ring mixer --> 25th chamber!
- Push ring mixer volume with air
- Left 8 chamber were empty, too many empty chamber...
Result
- qPCR result looked good.
- Most of chambers have 500pg~2400pg of amplicons.
- Left 8 chamber group were nearly empty, so that's why ch6 has almost no amplicon was detected.
- 25th chamber(Ring mixer) had less amount than other chamber did, but there was amplicon!
- It seemed that low conc. of ALS doesn’t work in the MDA device.
File:QPCR Result 2013-06-28 Edited.png File:QPCR Result MassTable 2013-06-28.png
Considerations
- In 20uL reaction, ALS is used 1.5uL for 1uL of cell, so 1.5/2.5 x 400mM KOH = 240mM in ALS+Cell while denaturing.
- Therefore current 400mM of KOH in ALS is much higher concentration for MDA device, because KOH conc. in ring mixer while denaturing is 357mM.
- Let’s try 240mM in ring mixer!
CoRE fragmentation
- Done @06/28