*After the excitation, the emission of fluorescent can last for a while even we turn off the exciataion. The colorful image still can be seen under the lamp light (white light). | *After the excitation, the emission of fluorescent can last for a while even we turn off the exciataion. The colorful image still can be seen under the lamp light (white light). | ||
==Suggestion== | ==Suggestion== |
Revision as of 00:19, 10 March 2009
E. coli labeling using SYBR Green I and YOYO-1
Objective
- To confirm the bacteria staining with revised procedures
- Using FITC filter, 20X objective, No cover slip
- The stained bacteria will be used for microwell application
Material and Samples
- Samples: Diluted E. coli, 10^6 CFU/uL (stored in PBS, 4C)
- YOYO®-1 iodide (491/509) - 1 mM solution in DMSO
- Diluted SYBR Green I (diluted in H2O, stored in 4C)
Procedures
- Drop 1 uL of E. coli on the glass slide
- Mix the E. coli with either 10 uL 2X SYBR Green I or 2 uL of YOYo-1 dye, gently pipetting for mixing
- Leave the slide in dark box (slide storage box)for incubation 15 min in room temp
Results
SNAP-151939-0309-sybr-20x-B&W Upper: SYBR Green I, 20X magnification, B&W SNAP-153406-0004-sybr-20x-fitc Middle: SYBR Green I, 20X magnification, FITC filter - Linear mode SNAP-153514-0005-sybr-20x-fitc Lower: SYBR Green I, 20X magnification, FITC filter - Best fit mode
- The YOYO-1 pictures were not available due to overexposure problem -> YOYO-1 dye needs to be diluted before use.
Discussion
- Both SYBR green dye and YOYO-1 dye can penetrate the bacteria and label on dsDNA without fixing bacteria
- Both SYBR green dye and YOYO-1 dye can been seen under FITC filter (filter designed for FITC dye)
- The SYBR green dye has fast photobleaching characteristics. Don't exposure to filtered light longer than 30 sec. The difference of emission reduced a lot after 30 sec.
- After the excitation, the emission of fluorescent can last for a while even we turn off the exciataion. The colorful image still can be seen under the lamp light (white light).
Suggestion
- Dilute the YOYO-1 dye before using -> find out best staining concentration.
- For the real application on microwell array, it needs to washout the unlabeled dye background before loading on the array.
- Ziess dye set list[1]
- Ziess filter set list[2]