Experiment 1: Cosmic 3760 probe preparation
Step 1: Real time PCR for cosmic probe

Reaction System                                        X8           X12
H2O                                     84.8ul      678.4 ul    1017.6 ul     
10xBuffer                                 10ul         80 ul       120 ul
SYBR  (50x)                              0.5ul          4 ul         6 ul 
*Mix the tube very well using vortex at this point to avoid locally concentrated
SYBR*        
dNTP(10mM)                                 4ul         32 ul        48 ul         
Primer Mix (100uM): Ap1V4IU +Ap2V4       0.4ul        3.2 ul       4.8 ul
template(cosmic3760)                     0.1ul        0.8 ul       1.2 ul 
Jumpstart Taq                            0.2ul        1.6 ul       2.4 ul
 Total                                    100ul        800 ul      1200 ul

Primer sequence: forward primer Ap1V4IU and reverse primer Ap2V4

Reaction program:

94c 2min -> 94c 30sec -> 60C 2min -> 72C 1min-> Plate read -> (94c 30sec -> 60C 2min -> 72C 1min)x 22 cycles -> 72C 5min -> 15C hold.

Note: 8 tubes were prepared and ran first to test if the reaction system is working properly.

Result: THe PCR reaction did not work as the curve never shot up. I am suspecting either the Taq enzyme or the 10x buffer is not working, so I will do a comparison to test the hypothesis. Since I ran a PCR reaction with Platium Taq and buffer few weeks ago, and it worked fine. So it will be used as reference.

Reaction System H2O 84.8ul 10x Platium Buffer 10ul SYBR (50x) 0.5ul

  • Mix the tube very well*

dNTP(10mM) 4ul Primer Mix (100uM): Ap1V4IU +Ap2V4 0.4ul template(cosmic3760) 0.1ul Jumpstart Taq 0.2ul

Total                                    100ul