Experiment 1: Cosmic 3760 probe preparation
Step 1: Real time PCR for cosmic probe

Reaction System                                        X8           X12
H2O                                     84.8ul      678.4 ul    1017.6 ul     
10xBuffer                                 10ul         80 ul       120 ul
SYBR  (50x)                              0.5ul          4 ul         6 ul 
*Mix the tube very well using vortex at this point to avoid locally concentrated
SYBR*        
dNTP(10mM)                                 4ul         32 ul        48 ul         
Primer Mix (100uM): Ap1V4IU +Ap2V4       0.4ul        3.2 ul       4.8 ul
template(cosmic3760)                     0.1ul        0.8 ul       1.2 ul 
Jumpstart Taq                            0.2ul        1.6 ul       2.4 ul
 Total                                    100ul        800 ul      1200 ul

Primer sequence: forward primer Ap1V4IU and reverse primer Ap2V4

Reaction program:

94c 2min -> 94c 30sec -> 60C 2min -> 72C 1min-> Plate read -> (94c 30sec -> 60C 2min -> 72C 1min)x 22 cycles -> 72C 5min -> 15C hold.

Note: 8 tubes were prepared and ran first to test if the reaction system is working properly.

Result: THe PCR reaction did not work as the curve never shot up. I am suspecting either the Taq enzyme or the 10x buffer is not working, so I will do a comparison to test the hypothesis. Since I ran a PCR reaction with Platium Taq and buffer few weeks ago, and it worked fine. So it will be used as reference.

Reaction Systems

reaction system 1: H2O 85.6ul 10xBuffer 10ul dNTP(10mM) 4ul Primer Mix (100uM): Ap1V4IU +Ap2V4 0.4ul SYBR (50x) 0.5ul Platium Taq 1ul template(cosmic3760) 0.1ul Total 100ul


reaction system 2: H2O 78ul 10x Platium Taq buffer 10ul Jumpstart Taq polymerase 0.2ul MgCl2(25mM) 6ul dNTP(10mM) 4ul primer mix (100uM) 0.4ul 50x SYBG I 0.4ul template(cosmic 3760) 1ul Total 100ul