MDA with PGP1F, 8th, Dev v9.1[edit]

  • Use new device (v9.1)


Experiment[edit]

KOH concentration for (ALS+Cell) mix[edit]
  • In 20uL reaction, the final concentration of KOH in ALS while denaturing is 240mM.
  • In the v7.3 version, 269mM of KOH in ALS buffer has used (v7.3 ring mixer device).
  • For this v9.1 device, 286mM of KOH in ALS buffer is used (Cell : (ALS-Cell) = 1 : 5.247).



  • Start : 8:30pm 07/17, End : 9:00am 07/18 --> about 13hr reaction
  • qPCR run for verification


Result[edit]

  • Amount of amplicons were good. (output channel 22 was clogged.)
  • Some of sample could be lost because of back flow as soon as opening valve
  • So the left 4 chamber were not filled well.

File:Backflow after opening valve 2013-07-18.png


  • I didn't run MDA in ALS chamber, but the reaction will be done in ALS chamber as well for the next experiment.
  • Also I'll try to push the (ALS+Cell) solution into compartment with H2O to prevent sample loss like this.