Qubit Fluorometer Protocol[edit]

Samples & Materials[edit]

  • Qubit fluorometer Invitrogen Cat. Q32857
  • Qubit Assay Tubes (500, Invitrogen Cat. no.Q32856) or Axygen PCR-05-C tubes (VWR, part number 10011-830)
  • Quant-IT DSDNA BR Assay Kit Invitrogen Cat. Q32850
  • Quant-IT DSDNA HS Assay Kit Invitrogen Cat. Q32851
  • Quant-IT RNA Assay Kit Invitrogen Cat. Q32852
  • Quant-IT Protein Assay Kit Invitrogen Cat. Q33211

File:Qubit1.jpg File:Qubit2.jpg

Calibrating the Qubit Fluorometer[edit]

Running a New Calibration[edit]

  • Insert standard #1
  • Press <Go>
  • Wait for ~5s
  • Insert standard #2
  • Press <Go>

Using the Last Calibration[edit]

  • Highlight <Use Last Calibration>
  • Press <Go>

Reading Samples[edit]

  • Insert sample
  • Press <Go>
  • Read the result on the display
  • Record the reading
  • Remove the sample form the instrument
  • Insert the next sample
  • Press <Go>
  • Repeat until all samples have been read

Calculation the Concentration of your Sample[edit]

  • Qubuit fluorometer gives values in either ug/ml or ng/ml
  • Concentration of your sample = QF value x (200/x)
*where QF = the value given by the Qubit fluorometer
        x = the number of microliters of sample you added to the assay tube
  • This equation generates a result with the same units as the value given by the Qubit fluorometer