MIDAS January 09, 2015[edit]

Overnight MDA 1/9 to 1/10

96 extractions 1/10, stored at 4 degC overnight

  • Xenoworks INJECT line flooded with solution, leading to errors in the remote controller.
  • Drierite also completely purple

Nextera and Ampure XP on 1/11

1st gel run on 1/12, random sample of 8 reactions. No positives. Large amount of high-mass material in each lane, including empty lanes. Likely DNA contamination in TBE buffer.

Remaining gels run 1/16 with fresh buffer. No positives

File:MidasGels150109Slide1.jpg File:MidasGels150109Slide2.jpg File:MidasGels150109Slide3.jpg File:MidasGels150109Slide4.jpg File:MidasGels150109Slide5.jpg

No positives, same as last experiment. Need to run positive controls next time.

Regenerated drierite (200 degC for 1 hour)