Amicon Ultra Centrifugal Filters[edit]
Amicon Ultra Centrifugal filters are microcentrifuge filters that allow you to purify and buffer exchange various MW cutoff biomacromolecules. They use a MW cutoff filter, say 10 kDa, to allow dialysate through while not passing the biomacromolecules. They can be used for buffer exchange via the following general protocol:
- CentrifugalFilters-Purification.png
Purification/Concentration
- CentrifugalFilters.png
Buffer Exchange
Protocol-Filtration[edit]
- Add X uL sample and buffer to 500 uL total volume
- Spin down at 14000 x g for appropriate time (15 minutes for 22 uL volume)
- Discard flow through tube and get a new one
- Invert filter inside flow through tube and spin down at 14000xg for 1 minute
- Alternately, extract sample using pipette
Protocol-Dialysis[edit]
- Add X uL sample and new buffer to 500 uL total volume
- Spin down at 14000 x g for appropriate time (15 minutes for 22 uL volume)
- Discard flow through and add 500 uL new buffer
- Repeat steps 2 and 3 as many times as desired
Dialysis Table[edit]
Starting Sample Volume | 500 uL dilution | Original Buffer Concentration | Centrifuge Time |
50 | 10 | 10% | 15 min |
20 | 25 | 0.4% | 15 min |
20 | 25 | 0.016% | 15 min |